Falsely undetectable TSH in a euthyroid patient

Abstract

Background

Measuring thyroid hormone levels is essential and helpful in the diagnosis of thyroid diseases.

Case presentation

We had a patient with undetectable serum TSH level by the Siemens immunoassay, whereas by ELISA and Roche-Abott immunoassay, serum level of TSH was in the normal range.

Conclusion

If the result of TSH level revealed very low or undetectable through one specific method, it should not be considered necessarily as a symptom of hyperthyroidism. It is to be requested to measure by the other methods with other devices too.

Background

Measuring thyroid hormone levels, especially TSH is essential and helpful in the diagnosis of thyroid dysfunction or diseases. Immunoassay is currently the best way to measure hormone levels. Of course, other methods, such as ELISA and gamma counter, are also used in some laboratories.

Case report

A 24-year-old woman attended to Al-Mahdi Clinic Lab with a request for thyroid function tests and other routine checkup tests. This patient had no history of disease or symptoms of malfunctioning of thyroid gland. Test results showed that total and free serum levels of T3 and T4 were normal, whereas TSH levels were less than 0.004 IU/ml (Normal range for adults 0.24–5.4 IU/ml). By confirming the results, Thyroid function tests were repeated by new sampling after 1 week and the same results were obtained. The method were used for detecting serum TSH level was immunoassay with Siemens Immulite 2000XPi device. The patient was referred to a physician for further evaluation and her physician again requested thyroid function tests with anti-TPO. With advice of his doctor, she was referred to another lab for doing these tests. The results of the second laboratory showed that all of Thyroid function tests were in normal range, so she returned to the first laboratory for consultation. The patient had no previous history of thyroid disease or cervical pain.

The case was re-sampled in the first laboratory and the TSH test result was less than 0.004 IU/ml as before. In the second laboratory the ELISA method was used for measuring serum TSH level. The first laboratory negotiated with the patient for assurance and her sample was sent to a third laboratory for measurement of serum TSH level by other device with Electro Immunoassay method. In the third lab, the Abbott Architect device was used, and again normal range of serum TSH level was taken (Table 1).

Table 1 Results of Serum TSH level with different devices and methods in three laboratories

Lab manager and physician in Al-Mahdi Clinic explained to the patient about the TSH variant in some people and she was assured that she was not hyperthyroid and doesn’t need any treatment.

Discussion & conclusion

Serum TSH test is routinely performed in medical laboratories to diagnose and evaluate thyroid gland in patients. Third-generation TSH is the most sensitive one. Immunoassay is currently the method of choice for these tests [1]. However, immunoassays are highly sensitive to variable influences that might lead to false results. False results can be due to the presence of macro-TSH, biotin, Antistreptavidin antibodies, Antiruthenium antibodies and Heterophilic antibodies [2, 3].

Manufacturers of immunoassays kits are trying to reduce the impact of such compounds by adding blocker to the solution or by notifying the user in the brochures. In these cases, clinical records, use of any multivitamin and other medications, history of blood transfusions, autoimmune diseases or direct contact to animals should be considered. When the patient in our study was referred to the laboratory, she had no major disease and did not take any multivitamin containing biotin or other drugs.

She had undetectable serum TSH with normal T3 and T4 levels so she was considered likely to have nodular goiter or other diseases resulting in hyperthyroidism but she had no history of any symptoms of these diseases. The presence of antistreptavidin antibodies, antiruthenium antibodies, and heterophilic antibodies was not considered because these antibodies could also interfere with the results of serum T3 and T4 levels [4].

However, there are other studies which show free T3, freeT4 and T4 levels could be in normal range with increased or decreased serum’s TSH level due to the presence of heterophilic antibodies [5, 6].

Due to the non-detectable TSH by the immunoassay with Siemens Immulite xp2000 and normal-level TSH reported by different other immunoassay devices, she could be considered a TSH variant case.

However, genetic testing should be performed to prove the mutation and these cases should also be investigated for the presence or absence of heterophilic antibodies. In our case, she didn’t consent to further evaluations and she agreed to perform diagnosis tests until her thyroid health was recovered.

Measurement of serum TSH by existing commercial methods of immunoassay is based on the sandwich method in which a robust antibody with a detectable antibody targets in two different regions of TSH. One antibody binds to the beta-TSH and the other one binds to the alpha-beta-TSH. Mutations in the beta-TSH region cause different results in the immunoassays of Siemens and other platforms and this mutation is benign. In our patient, findings were showed the normal bioactivity of variant TSH include clinical euthyroid, normal thyroid hormone levels and normal TSH in assays other than Siemens [7].

Drees JC and his colleagues identified 20 euthyroid patients (19 from South Asia and 1 Iranian) who had undetectable serum TSH level and they diagnosed hyperthyroidism and even began treatment for these patients. Subsequent reassessments revealed that the mutation in the beta-TSH (R55G) region, ie, the conversion of arginine to glycine amino acid, produced different false-negative results on serum TSH levels in different immunoassays. Arginine 55 is a part of the epitope that is identified in immunoassays and its mutation can explain the lack of TSH detection in immunoassays [8, 9].

When the result of serum TSH level is undetectable but the other thyroid function tests are normal range, the TSH test should be repeated by other different methods.

Availability of data and materials

Not applicable.

Abbreviations

Anti-TPO:

Thyroid-peroxidase antibodies

ELISA:

Enzyme-Linked Immunosorbent Assay

T3:

Triiodothyronine

T4:

Thyroxine

TSH:

Thyroid-stimulating hormone

References

  1. Edith T, Starich GH, Mazzaferri EL. Sensitivity, specificity, and cost effectiveness of the sensitive thyrotropin assay in the diagnosis of thyroid disease in ambulatory patients. Arch Intern Med. 1989;149(3):526–32.

    Article  Google Scholar 

  2. Piketty ML, Polak M, Flechtner I, Gonzales Briceno L, Souberbielle JC. False biochemichal diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences. Clin Chem Lab Med. 2017;55(6):780–8.

    Article  CAS  Google Scholar 

  3. Mogolu S, Armston AE, Mozley E, Nasruddin A. Heterophilic antibody interference affecting multiple hormone assays: is it due to rheumatoid factor? Scand J Clin Lab Invest. 2016;76(3):240–2.

    Article  Google Scholar 

  4. Favresse J, Burlacu MC, Maiter D, Gruson D. Interferences with thyroid function immunoassays: clinical implications and detection algorithm. Endocr Rev. 2018;39(5):830–50.

    Article  Google Scholar 

  5. Ismail AA, Walker L, Barth H, Lewandowski C, Jones R, Burr A. Wrong biochemistry results: two case reports and observational study in 5310 patients on potentially misleading thyroid-stimulating hormone and gonadotropin immunoassay results. J Clin Chem. 2002;48(11):2023–9.

    Article  CAS  Google Scholar 

  6. Lewandowski C, Dabrowska K, Lewinski A. Case report: When measured free T4 and free T3 may be misleading. Interference with free thyroid hormones measurements on Roche and Siemens platforms. J Thyroid Res. 2012;5(1):11.

    Article  Google Scholar 

  7. Pappa T, Johannesen J, Scherberg N, Torrent M, Dumitrescu A, Refetoff S. A TSHß variant with impaired immunoreactivity but intact biological activity and its clinical implications. Thyroid. 2015;25(8):869–78.

    Article  CAS  Google Scholar 

  8. Drees JC, Stone JA, Reamer CR, Arboleda VE, Huang K, Hrynkow J, et al. Falsely undetectable TSH in a cohort of south Asian euthyroid patients. J Clin Endocrinol Metabol. 2014;99(4):1171–9.

    Article  CAS  Google Scholar 

  9. Rahimkhani M, Einollahi N, Khavari Daneshvar H, Dashti N. Survey of serum procalcitonin in cirrhotic patients. Acta Medica Iranica. 2013;51(3):153–6.

    CAS  PubMed  Google Scholar 

Download references

Acknowledgments

We appreciate Mr. Alireza Khodabandeh who has helped us in this project.

Funding

There is no source of funding for the research reported.

Author information

Authors and Affiliations

Authors

Contributions

All authors have read and approved the final version of the manuscript for publication.

Corresponding author

Correspondence to Monireh Rahimkhani.

Ethics declarations

Ethics approval and consent to participate

The present case report was approved by local Tehran University ethic committee.

Consent for publication

Consent was obtained from the patient.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Rahimkhani, M., Kazemian, K. & Daryasari, R.R. Falsely undetectable TSH in a euthyroid patient. Thyroid Res 13, 2 (2020). https://doi.org/10.1186/s13044-020-0076-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13044-020-0076-y